Abstract

We compared the intra-operative diagnostic value of CDS vs IA–DSA in identifying major and minor technical defects after CEA. Between August 1997 and December 1998, 138 consecutive patients undergoing 141 carotid endarterectomies were intra-operatively investigated with colour duplex scan and intra-arterial digital subtraction angiography. Thirty-six (25.5%) technical defects were identified. Four (11.1%) major defects were detected by both methods and they were immediately corrected. Fifteen (41.6%) minor defects were detected by both methods, thirteen (36.1%) minor defects were detected by colour duplex but ignored by angiography. Angiography detected four (11.1%) kinkings missed with the colour duplex. The overall sensitivity of both methods for major defect was 100%. The sensitivity of colour duplex for minor defects was 87% vs 59% for angiography. On the basis of our study, colour duplex could be considered the choice method for quality control after carotid endarterectomy.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.