Abstract
Study of the intra observer and inter observer reliability of the pulsatility index, calculated from pulsed Doppler recordings of three fetal vessels. Flow velocity waveforms (FVW) were recorded from the umbilical artery, the fetal descending aorta and the fetal internal carotid artery. Intra-observer reliability was assessed in six fetuses; ten repeated measurements were performed by one observer. Inter-observer reliability was studied in 14 fetuses; two observers performed two repeated measurements in each fetus. A tertiary referral hospital. High risk pregnancies with a gestational age ranging from 29 to 42 weeks. Analysis of variance with repeated measurements and a graphical method were used for data analysis. Intra-observer repeatability was expressed as Intraclass Correlation Coefficient (IntraCC). Inter-observer agreement was expressed as Interclass Correlation Coefficient (InterCC). IntraCC for umbilical artery, descending aorta and internal carotid artery were 0.91, 0.78, and 0.54, respectively. InterCC values for these vessels were 0.39, 0.45 and 0.34, respectively. No systematic differences between the two observers except for the fetal descending aorta, were apparent. IntraCC decreased remarkable when fetuses with absent end diastolic velocities were excluded from the analysis. The pulsatility index (PI) used for fetal measurements has a poor reliability. This is of serious concern when clinical use of FVW measurements is considered as a diagnostic tool.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.