Abstract
The purpose of the study was to examine if the higher susceptibility at 3.0 Tesla (T) compared to 1.5 T will affect the contrast in MR imaging of the liver after application of superparamagnetic iron oxide particles (SPIO). The study was approved by our institutional review board and informed consent was obtained. Seventeen healthy volunteers were examined in a prospective, intra-individual comparative study within one day on a 1.5 T and a 3.0 T MRI system. T2 weighted TSE sequences were acquired after bolus injection of a SPIO contrast agent. Image contrast and signal to noise ratio (SNR) were compared between the field strengths. Image contrast was calculated between the liver tissue and the kidneys / spleen / muscles and fluids. The students'T-test was used for statistical analysis. No influence of the higher field strength could be observed on image contrast except for the liver / muscle contrast. This was due to a distinct SNR increase of the muscle tissue at 3.0 T as a result of their relaxation properties. The higher susceptibility at 3.0 T compared to 1.5 T does not translate into a stronger signal attenuation of the SPIO enhanced liver parenchyma.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.