Abstract

BackgroundAnatomic anomalies in the ascending aorta may impair the implantation and testing of cardiovascular devices in humans and animal models.Case presentationWe present the rare case of an intra-aortic band in a German Landrace pig. During terminal animal testing, the band hindered the implantation of a left ventricular assist device (LVAD) with transventricular outflow graft across the aortic valve. After lower partial sternotomy, epicardial echocardiography displayed an intraluminal echogenic structure at the sinotubular junction causing unspecific flow turbulences. Under cardiopulmonary bypass, coring of the left ventricular apex was performed. Due to strong resistance in the proximal aorta, accurate positioning of the transventricular LVAD outflow graft was impossible. After euthanasia, necropsy revealed a fibrous band located at the sinotubular junction, dividing the lumen of the ascending aorta.ConclusionsThe occurrence of an intra-aortic band represents an extremely rare case of a most likely congenital anomaly. Awareness of such anomalies is important for planning and performing animal testing. Perioperative echocardiography may help to either remove such anomalies or allow discontinuing the procedure prior to device implantation.