Abstract

AIM:The study intended to estimate the intra- and inter-model reliability and agreement between the HemoCue Models 201+ and 301 in a blood bank donor screening setting.MATERIALS AND METHODS:Venous blood samples from 115 blood donors were analyzed by two Hemocue 201+ instruments and two HemoCue 301 instruments. The Lin's concordance correlation coefficient (CCC) as well as the Bland–Altman limits of agreement was measured.RESULTS:The 95% lower confidence interval (CI) of all the CCCs had a value >0.90. The CCC between instruments belonging to the same model was compared to instruments belonging to different models. The mean of HemoCue 301 was higher than HemoCue 201+ by 3.04 g/L with the 95% CI of the differences (i.e., the limits of agreement) between the two models ranging from −3.35 to 9.44 g/L. The bias between two instruments was lesser, and the limits of agreement were even narrower between instruments belonging to the same model.CONCLUSION:The inter-instrument reliability and agreement are satisfactory in a blood bank setting.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.