Abstract

Oral mucosal moisture determined using oral moisture-checking devices is used as a mouth dryness evaluation method. Such devices are capable of evaluating the state of mouth dryness in a simple manner and have applicability in a wide range of subjects; however, their intra- and inter-investigator reliabilities have not yet been clarified. This study aims to investigate the intra- and inter-investigator reliabilities of measuring oral moisture using an oral moisture-checking device for a wide range of age groups. Intra- and inter-investigator reliabilities were investigated in 28 young subjects and 19 older subjects aged ≥65years. Three trained investigators independently measured oral mucosal moisture values using an oral moisture-checking device. Intra-investigator reliability was assessed using the coefficient of variation (CV) and intraclass correlation coefficient (ICC) (1.1), and inter-investigator reliability was assessed using ICC (2.1). Mean CV was 0.015 and 0.016, mean ICC (1.1) was 0.806 and 0.877, and ICC (2.1) was 0.873 and 0.829 in the young and older subjects, respectively. In young subjects, the mean values of ICC (1.1) and ICC (2.1) of the oral moisture-checking device were 0.806 and 0.873, respectively, whereas in older subjects, these values were 0.877 and 0.829, respectively. Thus, this confirms that the examination of oral mucosal moisture using the oral moisture-checking device has sufficient intra- and inter-investigator reliabilities for a wide range of age groups.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call