Abstract

The purpose of this study was to determine whether the thickness of the intimal hyperplasia (IH) layer that accumulates within Palmaz-Schatz stents is dependent on stent size. Intravascular ultrasound (IVUS) and quantitative angiographic (QCA) studies were performed after stent implantation and at follow-up (5.4 ± 3.8 months) in 161 patients with 177 lesions treated with 221 Palmaz-Schatz stents. Stent and lumen cross-sectional area (CSA) were measured. IH CSA and thickness at follow-up were calculated and compared with stent CSA and circumference. Maximum IH CSA and thickness were measured at the smallest follow-up lumen CSA; mean IH CSA and thickness was averaged over the length of the stent. Maximum IH CSA measured 4.8 ± 2.4 mm 2, and mean IH CSA measured 2.8 ± 2.2 mm 2. Maximum IH thickness (at the smallest follow-up lumen CSA) measured 0.60 ± 0.36 mm, and mean IH thickness (over the length of the stent) measured 0.30 ±0.19 mm. There was a weak, but significant correlation between mean and maximum IH CSA versus stent CSA (r = 0.215, p < 0.0001 and r = 0.355, p < 0.0001, respectively). However, there was no correlation between mean or maximum IH thickness versus stent CSA (r = 0.018, p = 0.643 and r = 0.056, p = 0.463, respectively) or stent circumference (r = 0.002, p = 0.956 and r = 0.069, p = 0.361, respectively). IH thickness was found to be independent of the stent size. This explains the known higher frequency of restenosis in smaller stents compared with larger stents.