Abstract

We present in vivo testing of a parallel transmit system intended for interventional MR-guided cardiac procedures. The parallel transmit system was connected in-line with a conventional 1.5 Tesla MRI system to transmit and receive on an 8-coil array. The system used a current sensor for real-time feedback to achieve real-time current control by determining coupling and null modes.Experiments were conducted on 4 Charmoise sheep weighing 33.9-45.0 kg with nitinol guidewires placed under X-ray fluoroscopy in the atrium or ventricle of the heart via the femoral vein.Heating tests were done in vivo and post-mortem with a high RF power imaging sequence using the coupling mode. Anatomical imaging was done using a combination of null modes optimized to produce a useable B1 field in the heart. Anatomical imaging produced cine images of the heart comparable in quality to imaging with the quad mode (all channels with the same amplitude and phase).Maximum observed temperature increases occurred when insulation was stripped from the wire tip. These were 4.1℃ and 0.4℃ for the coupling mode and null modes, respectively for the in vivo case; increasing to 6.0℃ and 1.3℃, respectively for the ex vivo case, because cooling from blood flow is removed. Heating < 0.1℃ was observed when insulation was not stripped from guidewire tips. In all tests, the parallel transmit system managed to reduce the temperature at the guidewire tip. We have demonstrated the first in vivo usage of an auxiliary parallel transmit system employing active feedback-based current control for interventional MRI with a conventional MRI scanner.

Highlights

  • Metallic interventional devices pose a risk in MRI because they are susceptible to RF-­induced currents, which can produce dangerous tissue heating.1–­3 visualizing standard guidewires with MRI is difficult

  • We have demonstrated the first in vivo usage of an auxiliary parallel transmit system employing active feedback-b­ ased current control for interventional MRI with a conventional MRI scanner

  • Operating at low power in a conventional MRI scanner can provide a safety margin but with tradeoffs in imaging performance required for visualization of both anatomy and guidewires.[6]

Read more

Summary

Introduction

Metallic interventional devices pose a risk in MRI because they are susceptible to RF-­induced currents, which can produce dangerous tissue heating.1–­3 visualizing standard guidewires with MRI is difficult. Together, these factors present substantial obstacles to MRI-g­ uided cardiovascular interventions. These factors present substantial obstacles to MRI-g­ uided cardiovascular interventions This is unfortunate considering that MRI provides improved soft-­tissue contrast and can provide functional measures such as flow and eliminates radiation dose.[4,5] Similar risks of RF-­induced heating are present for any elongated conductive device used in cardiac interventions, including guidewires, exchange wires, or braided catheters. A range of approaches is possible, including the use of susceptibility-­ related effects,[9,10] auxiliary contrast markers,[11,12] or some form of a local signal detector mounted on or integrated into the device.13–­15 In like manner, detecting signal at the device tip has been achieved by modifying guidewires with active components,[12,16,17] but these modification approaches require development of bespoke interventional devices

Methods
Discussion
Conclusion
Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call