Abstract

This study compares variability associated with two visual mammographic density (MD) assessment methods using two separate samples of radiologists. The image test-set comprised of images obtained from 20 women (age 42–89 years). The images were assessed for their MD by twenty American Board of Radiology (ABR) examiners and twenty-six radiologists registered with the Royal Australian and New Zealand College of Radiologists (RANZCR). Images were assessed using the same technology and conditions, however the ABR radiologists used the BI-RADS and the RANZCR radiologists used the RANZCR breast density synoptic. Both scales use a 4-point assessment. The images were then grouped as low- and high-density; low including BIRADS 1 and 2 or RANZCR 1 and 2 and high including BI-RADS 3 and 4 or RANZCR 3 and 4. Four-point BI-RADS and RANZCR showed no or negligible correlation (ρ=-0.029; p<0.859). The average inter-observer agreement on the BI-RADS scale had a Kappa of 0.565; [95% CI = 0.519 – 0.610], and ranged between 0.328–0.669 while the inter-observer agreement using the RANZCR scale had a Kappa of 0.360; [95% CI = 0.308 – 0.412] and a range of 0.078-0.499. Our findings show a wider range of inter-observer variability among RANZCR registered radiologists than the ABR examiners.