Abstract

The aim was to compare the reliability of cephalometric analysis using a smartphone-based application with conventional computer-based imaging software. Pre-treatment cephalometric radiographs of 50 subjects (26 males, 24 females; mean age, 19.2 years; ±4.2) were traced using the OneCeph® application and Dolphin imaging software®. Two independent observers identified seventeen landmarks and measured fourteen cephalometric measurements at an interval of. Interobserver and intraobserver reliability were evaluated using the intraclass correlation coefficient. Student's t-test was used to compare the means of two measurement methods for observer 1 and observer 2. Additionally, the time taken to complete the cephalometric measurements was also compared between the two methods. Good (ICC 0.75-0.90) to excellent (ICC 0.90-1.00) interobserver and intraobserver reliability was observed for all hard and soft tissue measurements with both methods. No significant differences were found between the two measurement methods for both observers (p<0.05). OneCeph application took significantly more time to complete the analysis than Dolphin imaging software (p<0.001). Cephalometric measurements made through a smartphone-based application showed good to excellent interobserver and intraobserver reliability and are comparable with the computer-based software. Therefore, it can be recommended for clinical use. The time taken to complete the cephalometric measurements was more with a smartphone-based application (OneCeph application) compared to computer-based software (Dolphin imaging software).