Abstract
A newly marketed rapid agglutination kit for the identification of Staphylococcus aureus, Slidex Staph Plus (bioMérieux), was compared to Staphaurex Plus (Murex Diagnostics) and Pastorex Staph-Plus (Sanofi Diagnostics Pasteur). The study took place in three clinical microbiology laboratories in three different European countries. A total of 892 staphylococcal isolates, including 278 methicillin-sensitive S. aureus (MSSA) isolates, 171 methicillin-resistant S. aureus (MRSA) isolates, and 443 coagulase-negative staphylococcal isolates, were analyzed. The sensitivities (MSSA/MRSA) and specificities, respectively, were 98. 2% (98.9%/97.1%) and 98.9% for Slidex Staph Plus, 98.2% (98.2%/98. 2%) and 96.2% for Staphaurex Plus, and 98.7% (98.6%/98.8%) and 95.7% for Pastorex Staph Plus. The specificity of the Slidex Staph Plus kit was statistically significantly higher than the specificities of Staphaurex Plus and Pastorex Staph-Plus. The Slidex Staph Plus is a very reliable test for the identification of S. aureus.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.