Abstract
The wavelengths of He-Ne lasers stabilized by saturated absorption in iodine and designed at BIPM, NPL, and PTB have been compared simultaneously by beat frequency measurements. The intention of the comparisons was to get experience about the reproducibility of differently designed lasers. All lasers were stabilized by third harmonic locking but with different electronic circuits. Among other effects, the reproducibility appears affected also by a modulation amplitude dependent shift of the laser frequency f. For a suitable range of modulation amplitudes the reproducibility of the four investigated lasers has been estimated to be ±4 × 10−11 f.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: IEEE Transactions on Instrumentation and Measurement
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.