Abstract

A secondary reference material for lipoprotein(a) is required to standardize the measurement of lipoprotein(a) in clinical laboratories worldwide. Towards this aim, the International Federation of Clinical Chemistry Working Group for the Standardization of Lipoprotein(a) Assays has initiated a standardization project involving a total of 33 diagnostic company and clinical chemistry laboratories from 12 countries. In Phase 1, the analytical performance of 40 lipoprotein(a) assay systems was evaluated by testing sera and manufactured lipoprotein(a) calibrator materials for precision, linearity, and parallelism. Twenty test systems were nonoptimized according to the results for a pooled serum, which tested nonlinear in 16 systems and imprecise in 4. Acceptable analytical properties and harmonization of lipoprotein(a) values were shown by some commercial calibrators, suggesting their possible use as reference materials. This study highlights the problems that currently occur for lipoprotein(a) measurement in existing assay systems.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.