Activity measurements of a solution of 75Se, supplied by the BIPM, have been carried out by 21 laboratories within the framework of an international comparison. Seven different methods were used. Details on source preparation, experimental facilities and counting data are reported. The measured activity–concentration values show a total spread of 6.62% before correction and 6.02% after correction for delayed events, with standard deviations of the unweighted means of 0.45% and 0.36%, respectively. The correction for delayed events was measured directly by four laboratories. Unfortunately no consensus on the activity value could be deduced from their results. The results of the comparison have been entered in the tables of the International Reference System (SIR) for γ-ray emitting radionuclides. The half-life of the metastable state was also determined by two laboratories and found to be in good agreement with the values found in the literature.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.