Abstract

Over the last decade, hemodialysis with enhanced internal filtration (IF) has been investigated as an alternative to conventional dialysis. Several factors affect IF, including the geometry and permeability of hollow-fiber dialyzers. Although various studies have been performed, the association between IF and membrane permeability has not been fully examined because of the difficulty in measuring IF. Therefore, in this study, we set up an experimental circuit and attempted to directly measure IF as well as membrane permeability in five dialyzers. In the circuit, we placed two dialyzers of the same type in series, and a special sampling port between them, thereby making it possible to determine IF by measuring the extent to which blood was concentrated between the two dialyzers. We showed that a significant amount of IF occurred in this tandem-dialyzer circuit, ranging from 23.5 to 100 ml/min, which increased linearly with increasing membrane permeability. We also showed that membrane permeability was reduced in the first dialyzer to a greater extent than in the second one after four hours of circulation, suggesting that filtration caused substantial membrane fouling. In this study we practically demonstrated that membrane permeability is highly relevant to the phenomenon of IF.