Abstract

Brain damage due to retraction was studied morphologically and physiologically in the dog brain. Two methods of retraction were compared using a retractor with a strain gauge: 1) continuous retraction; and 2) intermittent retraction. Total retraction time was 60 minutes for each method. Brain damage started to appear at a retraction force of 30 gm with continuous retraction, and increased in proportion to the force used. The power spectrum of the electrocorticogram showed full recovery after the release of retraction when the retraction force was less than 40 gm. With intermittent retraction, the damage was morphologically minimal with a retraction force of less than 50 gm, and recovery of the power spectrum of the electrocorticogram was prompt in comparison with continuous retraction. The results indicate the superiority of intermittent over continuous retraction.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.