Abstract

Common ultrasound transducers utilize epoxy for bonding the active piezoelectric to the rest of the acoustic stack. This paper explores some of the potential benefits of replacing this soft epoxy layer with a harder metallurgical bond. The influence of varying the bond layer thickness was evaluated for epoxy- and intermetallic bonding layers in an ultrasound transducer, as was the effect of voids and delamination that may occur during fabrication of intermetallic bonds. It was found that the intermetallic bond could be made much thicker than an epoxy bond without degrading the transducer performance, and the influence by thickness variations was less. Small voids in the intermetallic where shown to cause a downshift in resonance frequency, whereas regions with delamination will cause reflections that degrade the performance of the transducer.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.