Abstract

In wireless power transfer systems, the transmitting coil dimensions can substantially affect the transmission range and alignment sensitivity. We found that a transmitting coil with larger inner and outer diameter has a wider transmission range and lower alignment sensitivity. Thus, we developed a larger coil (24×30 cm2) designed to be embedded in the back of a vest to power DC pumps for artificial hearts or LVADs. To significantly reduce the required voltage, the coil was divided into 8 segments with resonant capacitors. The coil was operated at 6.78 MHz and evaluated with a 5.3-cm diameter receiving coil. A circuit model for the energy coupling coils was developed to predict the output power and efficiency. Having a coil separation of 7.7 cm, the output power and efficiency of the energy coupling coils are higher than 48 W and 80%, respectively. The system was experimentally tested with a DC pump, demonstrating that the proposed coil segmentation technique can significantly reduce the transmitter voltage to a safe level (∼10 Vrms).

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.