The results of the oxidation induction time (OIT) tests indicate that the determination of OIT involves a high degree of uncertainty in respect of the measured data, particularly for low OIT values. This would seem to show that OIT measurements have an extremely critical significance for quality control purposes or lifetime predictions of polyolefin parts. Determining oxidation induction temperature (OIT*) could therefore be a valuable alternative for less stabilized polyolefins (low OIT values) in particular. However, OIT* measurement clearly indicates that the ability to distinguish between different samples decreases drastically as OIT* data increase. Generally, the evaluated data can be used to estimate the results of in-house and external OIT/OIT* measurements and their compatibility. This report should therefore support day-to-day work in analytical laboratories where oxidation induction values are measured by DSC.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.