Abstract

Computerized potentiometric stripping analysis (PSA) was used for the determination of lead in wine samples. With this analytical method the sample preparation is limited to dilution with an appropriate supporting electrolyte (0.5 mol/l HCl and 40 mg/l Hg2+). Due to very high sensitivity the time needed for one determination can be reduced to 2 min. This method for lead determination in wine was compared with results from voltammetry and ETAAS from various laboratories participating in two BCR studies. With PSA it is possible to determine cadmium, lead and copper in wine after a single dilution with an appropriate supporting electrolyte. For other elements such as Bi, Tl, Hg, Sn, As, Zn determination procedures are under development.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.