Abstract

A set of test samples, all containing ultrasonically equivalent tissue-mimicking material, was produced and measurements of ultrasonic speed and ultrasonic attenuation coefficients were made at seven laboratories using various techniques. The ultrasonic speed values agree well with one another, having a spread of about 0.3 per cent; thus, speed values for tissue parenchyma appearing in the literature are likely to be accurate. Values of ultrasonic attenuation coefficients agree fairly well with one another, with differences between individual values and the group mean of generally less than 20 per cent of the group mean.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.