Abstract

A collaborative study was carried out according to internationally recognised guidelines in order to establish the performance characteristics of an LC/MS method for the determination of the feed additive semduramicin (SEM) in poultry feed at the level (20–25 mg kg−1) authorised within the European Union. Fifteen laboratories participated in the validation study, and all reported results. The content of SEM in the tested materials, provided as blind duplicates, ranged from 11.5 mg kg−1, which corresponds to half the mean authorised level, to 45.0 mg kg−1, which corresponds to twice the mean authorised level. All the materials were analysed by the participating laboratories using two different quantification approaches: standard addition and external standard calibration. The relative standard deviation of reproducibility (RSDR) for both quantification approaches varied from 8% to 18%, corresponding to HORRAT values ranging from 0.8 to 1.5, which were therefore in all cases below the critical value of 2.0. Consequently, the proposed analytical method and both quantification approaches can be considered to be fully validated and transferable to the control laboratories and applied for the determination of SEM in poultry compound feed at authorised level within the frame of official control. Further steps in the administrative procedure aiming to adopt the method as part of an ISO/CEN standard are currently ongoing.