Abstract
This paper reports the results of the second interlaboratory calibration experiments, which were carried out in two parts. In the first, five types of commercially available and economically cheap membranes were tested as the barrier against thoron for “radon-only” measurements. It was found that all the membranes tested are good barriers against thoron and can be used in the “radon-only” device. In the second, actual calibration experiments were performed with LR-115 type-II and CR-39 track-etch detectors for different radon measuring configurations in a radon exposure chamber.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.