Abstract

Stable oxygen isotope compositions (δ(18)O values) of two commercial and one synthesized silver orthophosphate reagents have been determined on the VSMOW scale. The analyses were carried out in three different laboratories: lab (1) applying off-line oxygen extraction in the form of CO(2) which was analyzed on a dual inlet and triple collector isotope ratio mass spectrometer, while labs (2) and (3) employed an isotope ratio mass spectrometer coupled to a high-temperature conversion/elemental analyzer (TC/EA) where Ag(3)PO(4) samples were analyzed as CO in continuous flow mode. The δ(18)O values for the proposed new comparison materials were linked to the generally accepted δ(18)O values for Vennemann's TU-1 and TU-2 standards as well as for Ag(3)PO(4) extracted from NBS120c. The weighted average δ(18)O(VSMOW) values for the new comparison materials UMCS-1, UMCS-2 and AGPO-SCRI were determined to be + 32.60 (± 0.12), + 19.40 (± 0.12) and + 14.58 (± 0.13)‰, respectively.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.