Abstract

The precision and bias of the coulometric Karl Fischer ASTM method D1533-00 have been assessed in a collaborative ASTM round robin program for a group of 34 laboratories. The test materials used in this study included water saturated 1-octanol (WSO), water saturated 1-butanol (WSB), and a series of new and used transformer oil samples. Fundamental systematic biases have been demonstrated in the accuracy of the measurement of water in the WSO, WSB, and transformer oil samples. The systematic bias in the measurement of the WSO and WSB standards indicates that for some laboratories either the instruments were not accurate or the quantity of the standard was not measured accurately. A second type of systematic bias consisted of measurement errors associated with the selection of the Karl Fischer solvent that was used with each instrument, and this was superimposed upon the error in the measurement of the water in the standards. Using the statistical calculation method ASTM D 6300 the repeatability and reproducibility for water in transformer oil were found to be 7 mg/kg and 14 mg/kg respectively. The method detection limit of water was 8 mg/kg oil. The method bias was estimated based on the National Institute of Standards and Technology (NIST) Standard Reference Material (SRM) 2890, WSO, since no suitable reference material for water in transformer oil was available for this study.