Abstract

ObjectivesWe describe a new method, FlowSpot, to assess CMV-specific T-cell response by quantification of interferon-gamma (IFN-γ). CMV-specific, T-cell-released IFN-γ was captured by flow beads and measured via flow cytometry. In the present study, we used FlowSpot to assess CMV-specific T-cell response in healthy individuals. The FlowSpot results were compared with those of serological analysis and enzyme-linked immunospot (ELISpot) assay. MethodsExperimental results and parameter analysis were investigated by using serological, ELISpot, and FlowSpot assays. ResultsThe levels of IFN-γ, which is released from CMV-specific T-cells, were measured, and the results and parameter analysis showed a good correlation between FlowSpot and ELISpot. However, FlowSpot was more sensitive and better reflected the strength of IFN-γ secretion than did ELISpot. ConclusionsCompared to ELISpot, FlowSpot has a high sensitivity and is cost and time effective. Thus, this method can be used in wider clinical and scientific applications.