Abstract
Sensors based on whispering gallery mode resonators can detect single nanoparticles and even single molecules. Particles attaching to the resonator induce a doublet in the transmission spectrum which provides a self-referenced detection signal. However, in practice this spectral feature is often obscured by the width of the resonance line which hides the doublet structure. This happens particularly in liquid environments that reduce the effective Q factor of the resonator. In this paper we demonstrate an interferometric setup that allows the direct detection of the hidden doublet and thus provides a pathway for developing practical sensor applications.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.