Abstract

Dual-comb spectroscopy (DCS), which uses two optical frequency combs (OFCs), requires an accurate knowledge of the mode number of each comb line to determine spectral features. We demonstrate a fast evaluation method of the absolute mode numbers of both OFCs used in DCS system. By measuring the interval between the peaks in the time-domain interferogram, it is possible to accurately determine the ratio of one OFC repetition frequency (frep) to the difference between the frep values of the two OFCs (Δfrep). The absolute mode numbers can then be straightforwardly calculated using this ratio. This method is applicable to a broad range of Δfrep values down to several Hz without any additional instruments. For instance, the minimum required measurement time is estimated to be about 1 s for Δfrep ≈ 5.6 Hz and frep ≈ 60 MHz. The optical frequencies of the absorption lines of acetylene gas obtained by DCS with our method of mode number determination shows good agreement with the data from the HITRAN database.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.