Abstract
The casual finding of disparate results in the determination of an analyte when it is measured at different dilutions on the same sample can lead us to suspect the presence of analytical interference. We present the study of 3 patients in whom a positive interference was proved on the immunoturbidimetric measurement of antistreptolysin-O due to the presence of a monoclonal immunoglobulin IgM Kappa.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.