Abstract

A simple and highly selective method was developed for the routine determination of trace or ultratrace amounts of gold and palladium in geological and metallurgical samples. The method uses flow injection on-line preconcentration and separation with determination by flame atomic absorption spectrometry. Au and Pd in the sample are adsorbed on a 2-mercaptopyrimidine chemically modified silica gel (MPMSG) packed microcolumn in a 0.50M HCl medium and then eluted with 0.5 or 1.0% thiourea solution. The eluates are introduced into the flame atomic absorption spectrometer directly. With the use of a 0.85 mL microcolumn (about 0.14 g MPMSG packed), the present system tolerated concentrations of common base metal ions up to 25.0 mg/mL and concentrations of anions up to 100.0 mg/mL when Au(III) at 0.100 microg/mL and Pd(II) at 0.200 microg/mL were preconcentrated for 60 s with a sample flow rate of 5.0 mL/min. The limits of detection were 3.1 ng/mL for Au(III) and 6.1 ng/mL for Pd(II) with relative standard deviations of < or = 2.5%. The analytical results obtained by the proposed method for geological and metallurgical samples were in good agreement with the certified values.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.