Abstract

An amperometric urea-sensing electrode was prepared by immobilizing a tri-enzyme system of pyruvate oxidase (PyOx), pyruvate kinase (PK) and urea amidolyase (UA) on a polydimethylsiloxane (PDMS)-coated electrode. UA catalyzes the adenosine-5′-triphosphate (ATP)-dependent hydrolysis of urea to generate adenosine-5′-diphosphate (ADP); PK, the phospho-transferring reaction between ADP and phosphoenolpyruvate (PEP) to generate pyruvic acid; PyOx, the oxidation of pyruvic acid with oxygen. The oxygen consumption could be monitored by using a PDMS-coated electrode without interference from the PyOx-reaction product, hydrogen peroxide. Thus, the concentration of urea (5 μM–0.35 mM) could be determined from the decrease in the cathodic current at −0.4 V versus Ag/AgCl in a test solution containing ATP and PEP. The cathodic detection with the tri-enzyme system provided the urea determination which was free from the error caused by coexisting species, such as acetaminophen, uric acid, phosphate in the sample to be measured.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.