Abstract
We examined the effect of fetal hemoglobin and labile glycosylated hemoglobin on a number of diverse methods used to measure glycosylated hemoglobin. Samples were supplemented with various amounts of cord blood to give proportions of fetal hemoglobin ranging from 1 to 20% of total hemoglobin concentration. Procedures in which the separation of hemoglobin A1 from the major hemoglobin A fraction is based on differences in ionic properties (cation-exchange chromatography and electrophoresis) are subject to interference by fetal hemoglobin, whereas procedures that base the quantitation on other properties (colorimetry and affinity column chromatography) are not. The same procedures that are affected by the presence of fetal hemoglobin are also subject to interference by labile glycosylated hemoglobin. We conclude that the affinity chromatographic and colorimetric methods may give a more nearly accurate determination of glycosylated hemoglobin.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.