Abstract

In the course of development and validation of a gas chromatography–mass spectrometry (GC–MS) method for ramipril and its biologically active metabolite ramiprilat, evidence was found for an unknown interfering metabolite. Sample treatment included isolation from plasma or urine by solid-phase extraction, methylation with trimethylsilyldiazomethane and acylation with trifluoroacetic anhydride (TFAA). When liquid chromatography was used to fractionate plasma extracts prior to derivatization, the alkyl, acyl-derivative of ramipril was obtained from two separate LC fractions. Electrospray ionization mass spectral data, together with circumstances for the derivatization, were consistent with the presence of an N-glucuronide of ramipril. Interference from the metabolite was eliminated by including a wash step after extraction/alkylation, prior to acylation. The final assay had a lower limit of quantification at 1.0 nmol/L and a linear range of 1–300 nmol/L. Intra- and inter-batch precision for ramipril and ramiprilat in plasma or urine were better than 10 and 5% at 2 and 80 nmol/L, respectively.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.