Abstract
Abstract Data are given for the rapid determination of calcium in biologic material by flame photometry. An "automatic background-subtraction," photometer added to the available instrumentation eliminated cation interference, while the use of a chelating agent (EDTA) prevented interference from anions (chiefly phosphate). The procedure shows excellent precision and accuracy in comparison with standard chemical procedures and in the analysis of control sera.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have