Abstract

We report on the fabrication and performance of a gel microfluidic chip interfaced to laser desorption/ionization (LDI) mass spectrometry with a time-of-flight mass analyzer. The chip was fabricated from poly(methylmethacrylate) with a poly(dimethyl siloxane) cover. Sodium dodecyl sulfate-polyacrylamide gel electrophoresis was performed in the channel of the microfluidic chip. After electrophoresis, the cover was removed and either the PDMS chip or the PMMA cover was mounted in a modified MALDI ion source for analysis. Ions were formed by irradiating the channel with 2.95 microm radiation from a pulsed optical parametric oscillator (OPO), which is coincident with IR absorption by N-H and O-H stretch of the gel components. No matrix was added. The microfluidic chip design allowed a decrease in the volume of material required for analysis over conventional gel slabs, thus enabling improvement in the detection limit to a pmol level, a three orders of magnitude improvement over previous studies in which desorption was achieved from an excised section of a conventional gel.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.