Abstract
The fabrication and testing of a minimally intrusive (2 mm high, 10 mm diameter) biomedical interface pressure sensor are described. Such sensors are needed for the implementation of improved safety features in the next generation of automated intravenous regional anaesthesia (IVRA) systems. The sensor utilizes a structured elastomer as a deflection element sandwiched between the plates of a parallel plate capacitor device. Simple mechanical modifications allow sensitivity and zero offset adjustment. The sensor is housed in a package machined from an engineering polymer. The device is easily calibrated using either a bench-top or an on-body calibration procedure. The device is particularly sensitive to cuff artefacts arising from variations in cuff-wrap tightness and folding of the cuff. As such, it offers some promise for detecting potential hazard conditions which can occur during conventional IVRA procedures. For the purpose of unit conversion, 1 Pa=1 N/m 2, 1 MPa=1 N/mm 2 and 40 kPa≈300 mmHg.
Published Version
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.