Abstract

The coupling of capillary electrophoresis and inductively coupled plasma (ICP) mass spectrometry is described using a new direct injection nebulizer designed for liquid flow-rates of 1–15 μl min −1 (aqueous solutions). The liquid flow-rate working range was found to depend on the nebulizer design. An extra flow of electrolyte was used to complete the electrical circuit. This design was found to work better than the conductive paint and crack hole designs. The CE–ICP-MS system was tested by analysing mixtures of alkali/alkaline earth metals and Cr(III)/Cr(VI), using electrokinetic injection. With ICP-MS detection an optimal internal standard, with the same electrophoretic mobility as the analyte, can be used.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.