Abstract
A novel interface circuit, which provides the accurate voltage output related to resistance change for three-wire resistance temperature detector has been presented. Moreover, the proposed interface circuit is able to eliminate the effect of lead wire resistances. To reduces the nonidealities of current conveyors, the proposed circuit based on operational conveyors. The validation affirms by comparing the theoretical analysis between the proposed circuit and basic configuration. The PSPICE simulations were conducted five different values of lead resistances when using with Pt100. The performance of the proposed circuit was investigated through simulation results.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.