Abstract

We describe a method for direct intercomparison of terahertz permittivities at 200 GHz obtained by a Vector Network Analyzer and a Time-Domain Spectrometer, whereby both instruments operate in their customary configurations, i.e., the VNA in waveguide and TDS in free-space. The method employs material that can be inserted into a waveguide for VNA measurements or contained in a cell for TDS measurements. The intercomparison experiments were performed using two materials: petroleum jelly and a mixture of petroleum jelly with carbon powder. The obtained values of complex permittivities were similar within the measurement uncertainty. An intercomparison between VNA and TDS measurements is of importance because the two modalities are customarily employed separately and require different approaches. Since material permittivities can and have been measured using either platform, it is necessary to ascertain that the obtained data is similar in both cases.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.