Abstract

We report the corrected intercomparison of the World Radiometer Reference (WRR) irradiance scale and the Electrotechnical Laboratory (ETL) spectral irradiance scale. In addition, we confirm the intercomparison precision using the test facility where the irradiance of ETL 500 W standard lamp can be measured directly with the cavity radiometer. The results showed that the irradiance based on the WRR scale was 0.5–0.7% lower than the one based on the ETL scale.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.