Abstract

This work is focused on studying the capabilities of different Monte Carlo tools to complement the Full Energy Peak Efficiency (FEPE) calibration procedures of the Environmental Radioactivity Laboratory of the Universitat Politècnica de València, UPV. In this frame, detailed models of a High Purity Germanium detector have been implemented using MCNP6 and GEANT4. Accordingly, an inter-comparison with experimental values ensures the model validation and improves the analysis of the physics phenomena involved. The validation of the models is determined by a quantitative comparison between the simulated and the measured efficiencies over the energy range under study (59.54–1836.01 keV). The results show discrepancies between both Monte Carlo tools for 139Ce, 88Y and 60Co as GEANT4 is able to simulate the coincidence summing effect of these radionuclides.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.