Abstract

Four different methods for measuring ultrafine particle size distributions in the 3–10-nm particle diameter range are compared and discussed. These methods all use an ultrafine condensation particle counter (TSI Inc. Model 3025 or its prototype) as the detector, but use different approaches to determine the size of the particles counted. Size classification was achieved using a Hauke Model VIE-06 differential mobility analyzer, a specially configured TSI Model 3040S diffusion battery, an ultrafine condensation particle counter with a variable condenser temperature, and an ultrafine condensation particle counter with a pulse height analyzer for signals produced by the optical detector. The response of these systems to ultrafine particles of known size and composition was studied during a workshop held in Lund, Sweden, during July 1991. After this workshop, measurements of ultrafine particles were made on the Swedish icebreaker Oden during the International Arctic Ocean Expedition 1991 (August 1, 1991 throu...