Abstract
Recently, a scientific comparison of flatness measuring instruments at European National Metrology Institutes (NMIs) was performed in the framework of EURAMET. The specimen was a well-polished optical surface with a maximum measurement aperture of 150 mm in diameter. Here, we present an evaluation concept, which allows the determination of a mean flatness map taking into account different lateral resolutions of the instruments and different orientations of the specimen during measurement. We found that all measurements are in agreement with the mean flatness map within the uncertainty intervals stated by the participants. The aim of this scientific comparison is to specify an appropriate operation and evaluation procedure for future comparisons.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.