Abstract

Histological investigation of a lesion induced by radiofrequency ablation (RFA) treatment provides ground-truth about the true lesion size, thus verifying the success or failure of the RFA treatment. This work presents a framework for registration of two-dimensional large-scale histological sections and three-dimensional CT data typically used to guide the RFA intervention. The focus is on the developed interactive methods for reconstruction of the histological volume data by fusion of histological and high-resolution CT (MicroCT) data and registration into CT data based on natural feature points. The framework is evaluated using RFA interventions in a porcine liver and applying medically relevant metrics. The results of registration are within clinically required precision targets; thus the developed methods are suitable for validation of the RFA treatment.