Abstract

Fiducial markers visible on both imaging scans and the patient are a convenient and accurate method to register the patients head in stereotactic space preceeding an interactive image directed neurosurgical procedure. We have been using the Laitinen non-invasive relocatable stereo-adapter to carry the fidcucial markers both during the imaging and patient registration process. Since the adapter can be accurately remounted the surgical procedure can take place at any time interval after the imaging study. To compare registration accuracy studies were performed using a phantom carrying both surface fiducial markers and fiducial markers mounted on the stereo adapter. We have found that total system accuracy using either surface or adapter mounted markers and an optical tracking system is in the range of 3-5 mm and is acceptable for a broad range of neurosurgical applications.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.