Abstract

This paper presents the details of a combined hardware/software system, which has been developed to provide haptic feedback for teleoperated laparoscopic surgical robots. Surgical instruments incorporating quantum tunnelling composite (QTC) force measuring sensors have been developed and mounted on a pair of Mitsubishi PA-10 industrial robots. Feedback forces are rendered on pseudo-surgical instruments based on a pair of PHANTOM Omni devices, which are also used to remotely manipulate the robotic arms. Measurements of the behaviour of the QTC sensors during a simulated teleoperated procedure are given. In addition, a method is proposed that can compensate for their non-linear characteristics in order to provide a ‘realistic feel’ to the surgeon through the haptic feedback channel. The paper concludes by explaining how the force feedback channel is combined with a visual feedback channel to enable a surgeon to perform a two-handed surgical procedure better on a remote patient by more accurately controlling a pair of robot arms via a computer network.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.