While transcatheter valve-in-valve (ViV) implantation may reduce the number of lifetime surgical reinterventions in some patients with congenital heart disease, in other patients the bioprosthetic valve ring can lead to patient-prosthesis mismatch due to reduction of internal diameter if a transcatheter valve is implanted. A 49-year-old woman with multiple prior cardiac surgeries presented with heart failure due to bioprosthetic tricuspid valve stenosis. Initial predilation valvuloplasty suggested conventional ViV implant would produce patient-prosthesis mismatch, therefore, we intentionally fractured the bioprosthetic valve ring to facilitate optimal hemodynamics with implantation of a 29 mm transcatheter valve.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.