Abstract

Megavoltage photon intensity-modulated radiation therapy (IMRT) is typically used in the treatment of prostate cancer at our institution. Approximately 1% to 2% of patients with prostate cancer have hip prostheses. The presence of the prosthesis usually complicates the planning process because of dose perturbation around the prosthesis, radiation attenuation through the prosthesis, and the introduction of computed tomography artifacts in the planning volume. In addition, hip prostheses are typically made of materials of high atomic number, which add uncertainty to the dosimetry of the prostate and critical organs in the planning volume. When the prosthesis is bilateral, treatment planning is further complicated because only a limited number of beam angles can be used to avoid the prostheses. In this case study, we will report the observed advantages of using noncoplanar beams in the delivery of IMRT to a prostate cancer patient with bilateral hip prostheses. The treatment was planned for 75.6 Gy using a 7-field coplanar approach and a noncoplanar arrangement, with all fields avoiding entrance though the prostheses. Our results indicate that, compared with the coplanar plan, the noncoplanar plan delivers the prescribed dose to the target with a slightly better conformality and sparing of rectal tissue versus the coplanar plan.