Abstract

A dosimetric comparison was made of Helical Tomotherapy (HT) and Rapid'Arc® (RA) in 115 patients with head and neck carcinoma included in a prospective and multicentric study. HT and RA provided highly conformal plans that easily complied with dose volume constraints for organs at risk. HT reduced high doses to the planning target volumes (PTVs) compared to RA and provided a more homogeneous dose distribution but with an increased Non Tumoral Integral Dose (NTID) than RA. However, the clinical consequences of these dosimetric advantages and disadvantages need further investigation.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.