Abstract

Two-photon fluorescence (TPF) imaging system with dual-axis galvo mirror as scanning unit is built. Based on the optical path of imaging system built, mechanism of how intensity distortion generating in dual-axis galvo mirror scanning TPF imaging system is theoretically analyzed. In experiment, TPF imaging is realized with Rhodamine B sample through detecting TPF signal generated from Rhodamine B molecule at excitation of broadband femtosecond pulses. The imaging result verifies the existence of intensity distortion with distribution agreeing well with the theoretical analyzation. To eliminate the intensity distortion in TPF imaging result, a mathematical method is proposed in which a calibration matrix is calculated, As the calibration matrix is applied to the distorted imaging result of Rhodamine B sample, intensity distortion is relieved effectively and precise imaging result is achieved.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.